Tests with a longer duration cause heat to travel further into the sample, and, therefore, the resulting effusivity measurement represents a larger volume of material, or a larger scale of scrutiny. It is important to balance the scale of scrutiny appropriately with the expectations of homogeneity and consistency in the material evaluated.
For powder samples, the effusivity varies with packing conditions because air has an effusivity of 5 Ws½/m2·K. As the material is compressed, air is removed and the effusivity value increases. After a user-defined nominal pressure of 15 Pa is applied for most materials, the effusivity measurement reaches a plateau when it is plotted against applied pressure. This is referred to as the threshold pressure and should be independently determined for materials that are tested offline. Alternately, if a sensor used for effusivity measurements weighs at least 500 g and has a surface area equal to or less than a 6.4-cm (2.5-inch) diameter, adequate pressure is supplied. During on-line testing, the column of powder above a sensor face provides adequate pressure if the sensor is at least 20 cm below the powder surface.
Although the scale of scrutiny is only 200 to 500 mg, in off-line blend evaluation, each sample should be 25 g or more to establish reproducible compression and consistency in sample treatment. It is recommended that samples be placed in a container whose inner diameter is in close proximity to the outer diameter of the sensor. This prevents additional movement and spreading of the sample and results in improved reproducibility. In the case of a 6.4-cm diameter sensor, a 250-mL beaker makes an excellent single, sample holder. On the other hand, the threshold pressure will not affect evaluation results in on-line blend evaluation as long as the sensors are placed at a level well below the fill line in order for the downward pressure of the powder above the sensor to generate the equivalent pressure (see Definition of Terms and Symbols).
Other Factors Affecting Effusivity Quantification
Moisture
Effusivity values are sensitive to the amount of moisture contained in a sample. It is, therefore, important to maintain consistent sample moisture during calibrations and sample testing, unless it is the moisture variations that are the desired measurement. Because water has an effusivity of 1600 Ws½/m2·K, the content of 1% water increases the effusivity of a typical powder by 3%.
Environmental Temperature
The variation of environmental temperature may jeopardize the target precision and accuracy (see Definition of Terms and Symbols). It is recommended that the variation of the environmental temperature should be less than ±1 K to achieve absolute effusivity measurements with accuracies of 1% to 2%. If the environmental temperature cannot be controlled, then it should be measured and recorded. The temperature data then provide a basis upon which a suitable correction for temperature effects can be formulated. Because the major application of effusivity is the relative effusivity or determination of effusivity plateaus rather than absolute effusivity (e.g., blend evaluation and drying), and each set of measurements is taken at the same environmental temperature, this factor typically has little or no affect on the final results.
Sample Thickness
Because heat is applied and detected on the same side of the material, if the sample is completely penetrated during a test, the material on the other side of the sample will be factored into the test results, producing a false representation of the effusivity value of the sample. Therefore, any off-line sample that is tested should be thicker than the depth of heat penetration during the test. In blend evaluation, a typical 2-second test penetrates into the powders approximately 0.5 mm, which is why a 25-g sample placed in a container with sides matching the sensor dimensions is recommended. In an on-line measurement, the thickness of sample has no significance, because the sample is semi-infinite and the heat never reaches the far side of the sample. The worst-case scenario for minimum sample thickness would be to use the longest test time of 10 seconds and a high effusivity material (crystalline material with high moisture). In this scenario, a minimum sample thickness of 5 mm would be required.
Sensor/Sample Interface Temperature
The differences in test results caused by variability in the starting temperature are not negligible. It is, therefore, critical to monitor the sensor/sample interface temperature and ensure that the temperature is the same at the start of each test. To achieve this, a cooling time may be used between two successive experiments (see Definition of Terms and Symbols). Otherwise, the temperature of the sensor/sample interface should be measured and recorded. The temperature data then provide a basis upon which a suitable correction for the effects of starting sensor/sample interface temperature can be formulated.
(In)homogeneity
Effusivity at the surface of a sample may or may not represent the effusivity of the bulk of the sample. The extent of regional differences in effusivity will depend on the homogeneity or inhomogeneity of the sample. As a result, differences in homogeneity or inhomogeneity can be detected by regional differences in effusivity values. In pharmaceutical analysis, this characteristic enables effusivity technology to be an efficient, effective tool for blend evaluation. The limiting factor is the range of effusivities of the components in the mixture. If all the materials had effusivities with a span of 10 Ws½/m2·K, the technique would not be able to differentiate between blended and unblended material because the relative standard deviation (RSD) between sensors at different locations in the powder bed would start at a maximum of 1.25%.
Power Supply
Accuracy and precision of the test results depend directly on the accuracy and precision of the power supply for heating elements. Therefore, it is desirable to use a high-precision, constant current and voltage power supply in the constant current and voltage configuration, respectively. Also, the power supply should be capable of delivering sufficient current to the heating elements such that the sensor generates a statistically significant temperature rise at the sensor/sample interface. The ideal temperature rise at the sensor/sample interface ranges from statistically significant to less than 5 K.