BRIEFING
á1092ñ The Dissolution Procedure: Development and Validation.
This new general information chapter previewed here is intended to address several issues. Aspects of method development are mentioned only superficially in the current general information chapter
In Vitro and In Vivo Evaluation of Dosage Forms á1088ñ. This previewed chapter goes into greater detail and gives guidance to the analyst on developing meaningful dissolution methods. Similarly, the general information chapter
Validation of Compendial Methods á1225ñ only touches on the special considerations for validation of dissolution testing, whereas this previewed new chapter provides a typical detailed step-by-step approach for designing and validating a dissolution test. Lastly, the previewed chapter gives guidance to the analyst on validation and the use of new technologies and equipment in dissolution testing. This new informational chapter is loosely based on a stimuli article, A New General Information Chapter on Dissolution by Gray VA, Brown CK, Dressman JB, and Leeson LJ that appeared in
PF 27(6) [Nov.Dec. 2001]. This article was revised on the basis of comments from industry, the FDA and other regulatory bodies, and the USP Expert Council and its Committees. Readers are encouraged to send their comments concerning the format and the contents of this previewed information chapter to William Brown.
(BPC: W. Brown ) RTS39647-1; 40232-2; 40485-1; 40486-1
GENERAL COMMENTS
The dissolution procedure relies on an apparatus, a medium, and a study design chosen to be transferable and rugged and to yield acceptable data. A discriminating procedure yields data that distinguishes important differences in components and composition and/or method of manufacture between dosage forms. The value of dissolution as a quality control tool for predicting in vivo performance of a drug product is significantly enhanced if an in vitroin vivo relationship (correlation or association) is established; but this relationship is not required (see FDA Guidances). Occasionally the in vitro dissolution test is found to be more sensitive and discriminating than the in vivo test. From a quality assurance point of view, a more discriminating dissolution method is preferred because the test will indicate possible changes in the quality of the product before in vivo performance is affected.
The discriminatory power of the dissolution method depends on the method's ability to detect changes in the drug product. Ideally, the dissolution test conditions would show discrimination for product changes that may affect biopharmaceutical product performance. However, unless an in vitroin vivo correlation exists for the product, variations in dissolution behavior may or may not reflect variations in the product's in vivo performance. To determine if a dissolution method can show discrimination for product changes, the method should be challenged. There are several ways to challenge the discriminatory power of the method, but the most common approach is to test formulations manufactured with different parameters. Manufacturing variables can come from many sources, e.g., the drug substance, the drug product's formulation, and/or the drug product's manufacturing process. For example, the ability of the method to detect changes in the drug substance can be evaluated by testing products with drug substances having different particle sizes, crystal habits, solvation, surface areas, or synthetic pathways. The method sensitivity to drug product formulation changes can be challenged by testing products with different amounts or ratios of excipients or by testing formulation changes, such as wet granulation versus dry blend, or intra- versus extra-granulation. The method sensitivity to manufacturing process variables such as lubrication blend time, compression force, addition order, drying methods, coating methods, and equipment capability or size can also be evaluated. As different formulations are made using some of the variables just mentioned; the dissolution data are examined. If the data show a measurable difference for the key variables, then the method may be considered a discriminating test for critical manufacturing variables.
With regard to stability, the dissolution test should appropriately reflect relevant changes in the biopharmaceutical performance caused by temperature, aging, humidity, and photosensitivity. A properly designed dissolution test does not generate highly variable dissolution results nor is it associated with significant analytical solution stability problems. If these issues arise, the dissolution test design should be evaluated and appropriately revised.
MEDIUM
Physical and chemical data for the drug substance and dosage unit should be determined before selecting the dissolution medium. Two key properties of the drug are the solubility and solution state stability of the drug as a function of the pH value. When adjusting the composition of the medium to generate sink conditions, the influence of surfactants, pH value, and buffers on the solubility and stability of the drug should be evaluated. Key properties of the dosage unit that may affect the dissolution, enteric-coating, modified-release mechanism, and disintegration rate are hardness, friability, presence of solubility enhancers, and the presence of other excipients.
Selection of the dissolution medium is based, in part, on the solubility data and the dose range in order to ensure that sink conditions are met. The term sink conditions is defined as the volume of medium at least greater than three times that required to form a saturated solution of drug substance. A medium that fails to provide sink conditions may be justifiable if it is shown to be more discriminating or if it provides reliable data that otherwise can only be obtained with the addition of surfactants.
Using an aqueous-organic solvent mixture as a dissolution medium is discouraged; however, if an in vitroin vivo relationship is demonstrated that cannot be accomplished with a pure aqueous medium, this type of medium may be acceptable.
Purified water is often used as the dissolution medium but is not ideal for several reasons: the quality of water can vary depending on the source of water; the pH value is inherently difficult to measure because the pH value can vary from day to day and may also change during the run depending on the active substance and excipients; and finally, the surface tension may also be variable and dependent on the excipients in the formulation. Despite these limitations, water is inexpensive, readily available, disposed of easily, ecologically acceptable, and suitable for products with a release rate independent of the pH value of the medium. Further, if water is named as the medium in an official test, it is not necessary to change to another medium unless there is a compelling reason.
The dissolution characteristics of an oral formulation are evaluated over the physiologic pH range of 1.2 to 6.8 (1.2 to 7.5 for modified-release formulations). During method development, it may be useful to measure the pH before and after a run to see if the pH changes during the test. Selection of the most appropriate conditions for routine testing is then based on discriminatory capability, ruggedness, stability of the analyte in the test medium, and, where possible, relevance to in vivo performance.
Listed below are typical media for a dissolution test. Buffers and acids are prepared as directed for Buffer Solutions and Volumetric Solutions, respectively, under Reagents, Indicators, and Solutions. Simulated gastric fluid and simulated intestinal fluid are prepared as directed for Test Solutions under Reagents, Indicators, and Solutions. The following list is not exhaustive and not in order of preference:
Hydrochloric acid (typically between 0.1 and 0.001 N)
Acetate buffer (with a pH between 4.1 and 5.5; 0.05 M) Phosphate buffer (with a pH between 5.8 and 8.0; 0.05 M)
Polysorbate 20, polysorbate 40, polysorbate 60, and polysorbate 80 solutions
Sodium lauryl sulfate solutions
Lauryldimethylamine oxide solutions
Bile salts solutions and/or lecithin
Combinations of surfactant and acids or buffers
Simulated gastric fluid with or without enzyme
Simulated intestinal fluid with or without enzyme
For very poorly soluble compounds, aqueous solutions may contain a percentage of a surfactant (e.g., sodium lauryl sulfate, polysorbate, or lauryldimethylamine oxide) that is used to enhance drug solubility. The need for surfactants and the concentrations used could be justified by showing profiles at several different concentrations. Surfactants can be used as either a wetting agent or, when the critical micelle concentration is reached, to solubilize the drug substance. The molarity of the buffers and acids used can influence the solubilizing effect, and this should be evaluated.
The Biopharmaceutics Classification System describes the classification of compounds according to solubility and permeability.1 Biorelevant medium is a term used to describe a medium that has some relevance to the in vivo dissolution conditions for the compound. Choice of a biorelevant medium is based on a mechanistic approach that considers the absorption site, if known, and whether the rate-limiting step to absorption is the dissolution or permeability of the compound. In some cases, the biorelevant medium will be different from the test conditions chosen for the regulatory test, and the time points are also likely to be different. If the compound dissolves quickly in the stomach and is highly permeable, gastric emptying time may be the rate-limiting step to absorption. In this case, the dissolution test is to demonstrate that the drug is released quickly under typical gastric (acidic) conditions. On the other hand, if dissolution occurs primarily in the intestinal tract (e.g., for a poorly soluble, weak acid), a higher pH range (e.g., simulated intestinal fluid with a pH of 6.8) will be more appropriate. The fed and fasted state may also have significant effects on the absorption or solubility of a compound. Compositions of media that simulate the fed and fasted states can be found in the literature. These media reflect changes in the pH, bile concentrations, and osmolarity after meal intake, and, therefore, have a different composition than that of typical compendial media. They are primarily used to establish in vitroin vivo correlations during formulation development and to assess potential food effects,but are not intended for quality control purposes. For quality control purposes, the substitution of natural surfactants (bile components) with appropriate synthetic surfactants is permitted and encouraged because of the expense of the natural substances and the labor-intensive preparation of the biorelevant media.
The use of an enzyme in the dissolution medium is permitted (see Dissolution á711ñ) when dissolution failures occur due to pellicle formation with gelatin capsules or gelatin-coated products.
Normally for the basket and paddle apparatus, the volume of the dissolution medium is 500 mL to 1000 mL, with 900 mL as the most common volume. The volume can be raised to between 2 and 4 liters, depending on the concentration and sink conditions of the drug, but proper justification is expected.
The significance of deaeration of the medium may be determined: air bubbles can interfere with the test results, acting as a barrier to dissolution if present on the dosage unit or basket mesh; also, bubbles can cause particles to cling to the apparatus and vessel walls. On the other hand, bubbles on the dosage unit may increase the buoyancy and lead to an increase in the dissolution rate, or, decrease the dissolution rate by decreasing the available surface area. A deaeration method is described as a footnote in Dissolution á711ñ. Typical steps are to heat the medium, filter, and draw a vacuum for a short period of time. Other validated methods of deaeration are available and in routine use throughout the industry. Media containing surfactants are not usually deaerated because of excessive foaming. To determine if deaeration of the medium is necessary, dissolution run in nondeaerated medium versus deaerated medium should be performed.
STUDY DESIGN
Time Points
For immediate-release dosage forms, the duration of the procedure is typically 30 to 60 minutes, with a single time point specification that is adequate in most cases for Pharmacopeial purposes. Industrial and regulatory concepts of product comparability and performance may require additional time points, and this may also be a feature required for product registration or approval. Enough time points are to be selected to adequately characterize the ascending and plateau phases of the dissolution curve. According to the Biopharmaceutics Classification System referred to in several FDA guidances, highly soluble, highly permeable drugs formulated with rapidly dissolving products need not be subjected to a profile comparison if they can be shown to release 85% or more of the active drug substance within 15 minutes. For these types of products a one-point test will suffice. However, most products do not fall into this category. Dissolution profiles of immediate-release products not meeting the highly soluble/highly permeable/rapidly dissolving criteria typically show a gradual increase reaching between 85% and 100% at around 30 to 45 minutes. Thus, dissolution time points in the range from 15 (or 20), 30, 45, and 60 minutes are usual for most immediate-release product. Useful information may be obtained from other points, e.g., if dissolution is very rapid, 5 to 10 minutes to obtain more data or for slower-dissolving drugs, additional points after 60 minutes. Dissolution test times for compendial tests are usually then established on the basis of an evaluation of the dissolution profile data.
So-called infinity points can be useful during development studies. To obtain an infinity point, the paddle or basket speed is increased at the end of the run to at least 150 rpm for 30 to 60 minutes, after which time an additional sample is taken. Although there is no requirement for 100% dissolution in the profile, the infinity point can provide data that may supplement content uniformity data and may provide useful information about the formulation characteristics during the initial development.
For an extended-release dosage form, at least three test time points are chosen to characterize the in vitro drug release profile for Pharmacopeial purposes. Additional sampling times may be required for drug approval purposes. An early time point, usually 1 to 2 hours, is chosen to show that there is little probability of dose dumping. An intermediate time point is chosen to define the in vitro release profile of the dosage form, and a final time point is chosen to show essentially complete release of the drug. Test times and specifications are usually established on the basis of an evaluation of drug release profile data. For products containing more than a single active ingredient, drug release is determined for each active ingredient. Extended-release specifications are addressed in the general information chapter In Vitro and In Vivo Evaluation of Dosage Forms á1088ñ as well as other sources.2
Observations
Visual observations and recordings of the product dissolution and disintegration behavior are very useful because dissolution and disintegration patterns can be indicative of formulation or manufacturing process variables. To accomplish visual observation, proper lighting (with appropriate consideration of photodegradation) of the vessel contents and clear visibility in the bath is essential. Documenting observations by drawing sketches and taking photographs or videos is very instructive and helpful for those who are not able to observe the real time dissolution test. Observations are especially useful during method development and formulation optimization. Examples of typical observations include, but are not limited to, the following:
-
Uneven distribution of particles throughout the vessel, which could occur if particles cling to the sides of the vessels, if there is coning or mounding directly under the apparatus, if particles float at the surface of the medium, if film-coated tablets stick to the vessel, and/or if off-center mounds are formed;
-
Air bubbles on the inside of the vessel or on the apparatus or dosage unit. Sheen on the apparatus is also a sign of air bubbles. This observation would typically be made when assessing the need to deaerate the medium;
-
Dancing or spinning of the dosage unit, or the dosage unit being hit by the paddle;
-
Adhesion of particles to the paddle or inside of the basket, which may be observed upon removal of the stirring device at the end of the run;
-
Pellicles or analogous formations, such as a transparent sacs that are rubbery, swollen masses surrounding the capsule contents;
-
Presence of large floating particles or chunks of the dosage unit;
-
Observation of the disintegration rate (e.g., percent reduction in size of the dosage unit within a certain time frame);
-
For modified or enteric-coated products, complex disintegration of the coating, e.g., the partial opening and splitting apart (like a clamshell) or incomplete opening of the shell accompanied by the release of air bubbles and excipients.
Sampling
The disturbance of hydrodynamics of the vessel by sampling probes should be considered and adequate validation performed to ensure that the probes are not introducing a significant change in the dissolution rate.
Manual
Manual sampling uses plastic or glass syringes, a stainless steel cannula that is usually curved to allow for vessel sampling, a filter, and/or a filter holder. The sampling site must conform to specifications under Dissolution á711ñ.
Autosampling
Autosampling is a useful alternative to manual sampling, especially if the test includes several time points. However, since regulatory labs may perform the dissolution test using manual sampling, autosampling requires validation with manual sampling.
There are many brands of autosamplers, including semi- and fully-automated systems. Routine performance checks, cleaning, and maintenance as described in the pertinent standard operating procedures or metrology documents are useful for reliable operation of these devices.
Some instruments are equipped with sampling through the basket or paddle shaft. Proper validation (e.g., demonstrated equivalence to results with the usual sampling procedure) may be required.
Manual versus automated validation can be done either one of two ways. When the drug dissolution results are not highly variable, two concurrent runs (same sampling intervals, n = 6) using manual and automated sampling methods are compared using the criteria under Intermediate Precision. If the dissolution results are highly variable (i.e., the RSD is above 20% in time points at 10 minutes or earlier, and at or above 10% RSD in later time points), the analysis may be performed by pulling the sample from the vessel simultaneously by manual and automated sampling methods for each time point. Note that the correction for the volume withdrawn from the medium is doubled in this case.
If the results are not acceptable for the manual sampling, a discard step may be required or the discard volume may be increased. For the autosampler, a longer priming time may be needed if unacceptable results are obtained. If these changes do not work, then a change to another filter type may be needed.
Other aspects of automated validation may include the carryover of residual drug, the effect of an in-residence probe (simultaneous sampling as mentioned above may not be suitable in this case), adsorption of drug, and cleaning and/or rinse cycles.
Filters
Filtration of the dissolution samples is usually necessary to prevent undissolved drug particles from entering the analytical sample and further dissolving. Also, filtration removes insoluble excipients that may otherwise cause high background or turbidity. Prewetting of the filter with the medium is usually necessary.
Filters can be in-line or at the end of the sampling probe, or both. The pore size can range from 0.45 µm to 70 µm. The usual types are depth, disk, or flow-through filters. However, if the excipient interference is high, if the filtrate has a cloudy appearance, or if the filter becomes clogged, an alternative type of filter or pore size should be evaluated.
Adsorption of the drug(s) onto the filter is also evaluated. If drug absorption occurs, it may be necessary to increase the amount of initial filtrate discarded. If results are still unsuitable, an alternative filter material may be sought.
Filter validation may be accomplished by preparing a 100% standard solution and a completely dissolved sample solution (e.g., prepared as a typical sample in a vessel or a sample put in a beaker and stirred with a magnetic stirrer for 1 hour). The solutions are sampled (n = 3) by both the selected autosampler and the manual technique. Both sampling techniques may use the same or equivalent type of filter. The autosampler is programmed for a typical sample pull, and the manual method withdraws an aliquot from the solutions. The standard solution results are compared to an unfiltered standard. The analyst may compare the sample solution results to the same sample solution unfiltered and centrifuged. For the filter to be acceptable for use, the results of the filtered portions are to approach (within 98% to 102%) the original concentration of the unfiltered standard solution and the centrifuged sample solution.
Centrifugation
Centrifugation of samples is not preferred, as dissolution can continue to occur, and there may be a concentration gradient in the supernatant. A possible exception might be compounds that adsorb to all common filters.
ASSAY
The usual assay for a dissolution sample is either by spectrophotometry or high-pressure liquid chromatography (HPLC). The preferred method of analysis is spectrophotometry because results can be obtained faster, the analysis is simpler, and fewer solvents are used. HPLC methods are used when there is significant interference from excipients or among drugs in the formulation and/or to improve analytical sensitivity. It may be useful to have data for the drug obtained with a stability-indicating assay (e.g., HPLC chromatograms) in the medium of choice, even if the primary assay is based on a spectrophotometric method.
The validation topics described in this section are typical but not all-inclusive. The acceptance criteria are guidelines only and may differ for some products. Other considerations may be important for special dosage forms. The extent of validation depends on the phase of the product development. Full validation takes place at or just before Phase III clinical studies. For profile testing, validation studies may be necessary at each relevant time point. For products containing more than a single active ingredient, the dissolution method should be validated for each active ingredient.
Specificity/Placebo Interference
The placebo consists of all the excipients and coatings (inks, sinker, and, when appropriate, the capsule shell) without the active ingredient. Placebo interference may be determined by weighing three samples each of the placebo blend, equal to or greater than the highest and lowest strength. Transfer the samples to separate vessels filled with dissolution medium at 37

. Stir for around 30 minutes to 1 hour at 150 rpm using the method apparatus. Observe whether the blend becomes wetted. Analyze and calculate the percent interference at each strength (n = 3) by comparing it to the 100% standard of that strength (or optionally using the selected
Q value) by the formula:
100C(AP / AS)(V/L),
in which C is the concentration, in mg per mL, of the standard; AP and AS are the absorbances of the placebo and the standard, respectively; V is the volume, in mL, of the medium; and L is the label claim, in mg. The mean is determined, and the interference does not exceed 2%.
NOTEFor extended-release products, a placebo version of the finished dosage form may be more appropriate to use than blends because this placebo formulation will release the various excipients in a manner more nearly reflecting the product than a simple blend of the excipients. The placebo formulation is necessary particularly if one of the released excipients does influence the baseline of the analytical method. For extended-release products, the interference must be determined at each sampling point in the release profile. If the placebo interference exceeds 2%, then HPLC analysis is needed; alternatively, another wavelength, algorithm analysis, or a baseline subtraction (correction factor) may be feasible.
Spectrophotometric Analysis
Analyses may be performed with spectrophotometers that have diode array or variable wavelength detectors. Samples may be automatically introduced into the spectrophotometer using autosippers and flow cells. Routine performance checks, cleaning, and maintenance as described in the standard operating procedures or metrology documents are useful for reliable operation of these instruments. Flow cells ranging from 0.02 to 1 cm are typically used. Cell alignment and air bubbles could be sources of error. The smaller path length flow cells are used to avoid diluting the sample; however, acceptable linearity and standard error should be demonstrated.
During analysis, standard solutions are typically prepared and analyzed at just one concentration at 100% of the dosage strength (or the selected Q value). During profile analysis, other concentrations may be useful. The recorded standards are aliquots from the same standard solution. A typical blank, standard, and sample may be analyzed in a sequence that brackets the sample with standards and blanks, especially at the beginning and end of the analysis. In most cases, the mean absorbance of the dissolution medium blank may not exceed 1% of the standard. Values higher than 1% must be evaluated on a case-by-case basis. The typical relative standard deviation (RSD) for UV analysis is usually £ 2%.
The absorptivity is calculated by dividing the mean standard absorbance by the concentration, in mg per mL, divided by the flow-cell path length in cm. After enough historical data is accumulated, an acceptable absorptivity range for the analyte (using the appropriate flow cell) may be determined. This value may be useful in troubleshooting aberrant data. Fiber optics as a sampling and determinative method, with proper validation, is an option.
Examine the UV spectrum of the drug in solution to select the wavelength for maximum absorbency. In some cases, there may be other wavelengths chosen, as with measuring degradants along with the intact drug substance.
HPLC
For HPLC analysis, the compatibility of dissolution media and mobile phase may be examined, especially if large injector volumes (over 100 µL) are needed. Samples are normally analyzed with HPLC using a spectrophotometric detector and an auto-injector. Single injections of each vessel time point with standards throughout the run constitute a typical run design. System suitability tests include, at a minimum, the retention time window and precision. Typically, the system suitability of an HPLC analysis is less than or equal to 2% RSD for five or six standard determinations. The standard level is typically at the 100% label claim level, especially for a single-point analysis.
Preparation of the placebo samples for the HPLC analysis is performed in the same way as the spectrophotometric analysis. Look for peaks eluting at the same retention time as the drug. If there are extraneous peaks, inject the standard solution, and compare retention times. If the retention times are too close, spike the placebo solution with the drug. Chromatograms may also be obtained over an extended run time using the blank (dissolution medium), standard, and sample solution to identify late eluters that may interfere with subsequent analyses.
If possible, the validation documentation may include overlaid representative chromatograms or spectra of blank dissolution medium, a filtered placebo solution, a standard solution, and a filtered dissolution sample. Absence of interfering peaks in the placebo chromatogram or lack of absorbance by the placebo at the analytical wavelength demonstrates specificity.
Linearity and Range
Organic solvents may be used to enhance drug solubility for the preparation of the standard solutions; however, no more than 5% (v/v) of organic solvent in the final solution should be used, unless validated.
Linearity and range are typically established by preparing five standard solutions of the drug, ranging in concentration from approximately ±20% below the lowest expected concentration to ±20% above the highest concentration during release from the specific dosage unit strength. As further dosage strengths are added, the same testing scheme applies, unless some concentrations have already been tested. For spectrophotometric analysis, this scheme may be altered if different flow-cell sizes are used.
All solutions are made from a common stock if possible. The diluted solutions may be read, at least, in duplicate (two test tubes of the same solution) with spectrophotometric analysis and two injections of the same solution in different vials for HPLC analysis. For the highest concentration, the absorbance values may not exceed the linearity limits of the instrument.
Linearity is typically calculated by using an appropriate least-squares regression program. Typically, a square of the correlation coefficient (r2 ³ 0.98) demonstrates linearity. In addition, the y-intercept must not be significantly different from zero at the 95% confidence limit.
Accuracy/Recovery
For full validation, drug powder at ±20% of the lowest expected concentration to ±20% of the highest concentration during release from the specific dosage unit strength is typically used. A minimum of three concentration levels is evaluated. The capsule shell, coating blend, inks, and sinker are also added where appropriate. Each level is tested as n = 3. The amount of placebo blend weighed for each level is the same as the total excipient weight of each tablet or capsule for the dosage strength being validated. For Apparatus 1 and Apparatus 2, the mixture of excipients and drug powder may be tested according to the conditions specified in the method.
In cases of poor drug solubility, if feasible, the stock solution may be prepared by dissolving the drug substance in a small amount of organic solvent and diluting to the final concentration with dissolution medium. An amount of stock solution equivalent to the targeted label claim may be added to the vessel instead of drug powder. [NOTEThe solutions may be made in volumetric flasks instead of vessels.
] The percentage of organic solvent should be minimized (ideally £5% of the final medium composition) as it may change the solubility of the excipients and bias the results.
The measured recovery is typically 95% to 105% of the weighed amounts. Bracketing or matrixing of multiple strengths may be useful.
A special case for validation is the acid stage (see Delayed-Release (Enteric-Coated) ArticlesGeneral Drug Release Standard under Drug Release á724ñ. The limit of not more than 10% should be validated. If the compound degrades in acid, the validation of the method is complicated but usually can be performed using the degradant, especially if 100% rapid degradation in-situ occurs.
Precision
REPEATABILITY
Repeatability is determined by replicate measurements of a standard solution. It can be measured by calculating the RSD of the multiple injections, by spectrophotometric readings for each standard solution, or from the accuracy or linearity data.
INTERMEDIATE PRECISION
Intermediate precision may be evaluated to determine the effects of random events on the precision of the analytical procedure. The precision can be across the range of product strengths. Typical variations to study include days, analysts, and equipment. The use of an experimental matrix design is encouraged for evaluation of intermediate precision. If possible, the intermediate precision can be performed using a well-characterized lot of drug product of tight content uniformity. In cases where this well-characterized product is not available, placebo and active ingredient may be used to identify intermediate precision.
The dissolution profiles on the same sample (run twice, n = 12) may be run by at least two different analysts, with each analyst preparing the standard solutions and the medium. The analysts typically use different dissolution baths, spectrophotometers or HPLC equipment (including columns), and autosamplers and perform the test on different days. This procedure may not be necessary to perform for each strength; instead bracketing with high and low strengths may be acceptable.
Typical acceptance criteria for this type of precision are that the difference in the mean value between the dissolution results from the first analyst to the second analyst does not exceed an absolute 10% at time points with less than 85% dissolved and does not exceed 5% for those remaining time points above 85%. Acceptance criteria may be product specific, and other statistical tests and limits may be used.
Robustness
The evaluation of robustness may be considered during the development phase but more typically is done at the time of full validation. This assesses the impact of making small, deliberate changes to the dissolution conditions. Experiments to test for robustness can use n = 3 if the product exhibits normal variability and n = 6 for highly variable products (i.e., the RSD is above 20% in time points at 10 minutes or earlier, and 10% or above RSD in later time points).
One run each may be performed using 90%, 100%, and 110% of the medium concentration (e.g., surfactant, w/v %) in the method. For buffered media, runs of 0.5 pH units above and below the method pH can be performed. The buffer capacity may be varied by changing the total buffer concentration while keeping the relative amounts of each buffer component the same (see above for suitable variations).
HPLC includes variations of the mobile phase composition, flow rate, pH changes, and column type, brand, lot, or age. System suitability criteria assess acceptability of the method. For a more in-depth discussion of robustness and other validation topics for HPLC, refer to the general information chapter Validation of Compendial Methods á1225ñ. UV could include a variation of wavelength of ±2 nm, depending on the absorbance profile.
Standard and Sample Solution Stability
The standard solution is stored under conditions that ensure stability. The stability of the standard is analyzed over a specified period of time, using a freshly prepared standard solution at each time interval for comparison. The acceptable range for standard solution stability is typically between 98% and 102%.
The sample solution is stored at room temperature in a glass test tube wrapped securely in paraffin or in a capped vial. The sample is analyzed over a specified period of time using the original sample solution response for comparison. The typical acceptable range for sample solution stability may be between 98% and 102% when compared to the initial analysis of the sample solutions. If the solution is not stable, aspects to consider could be temperature (refrigeration needed), light protection, or container material (plastic or glass).
The method may state that the standards and samples should be analyzed within the time period demonstrating acceptable standard and sample solution stability.
ACCEPTANCE CRITERIA
Typical acceptance criteria for the amount of active ingredient dissolved, expressed as a percentage of the labeled content
(Q), are in the range of 75% to 80% dissolved. A
Q value in excess of 80% is not generally used because allowance should be made for assay and content uniformity ranges.
3 Acceptance criteria including test times are usually established on the basis of an evaluation of the dissolution profile data.
High variability in results can make it difficult to identify trends or effects of formulation changes. Dissolution results may be considered highly variable if the RSD is greater than 20% at timepoints 10 minutes or earlier and greater than 10% RSD in later time points. The source of the variability could be investigated, when practical, and one should strive to reduce variability whenever possible. The two most likely causes are the formulation itself or mechanical artifacts associated with the test procedure (e.g., coning or tablets sticking to the vessel wall).
Occurrences such as those mentioned in the previous section on observations can be a strong indication that the dissolution test itself is creating variability. Any time the dosage contents do not disperse freely throughout the vessel in a uniform fashion, aberrant test results could occur. Depending on the problem, the usual remedies include changing the apparatus type, speed of agitation (flowrate), deaeration, consideration and/or examination of sinker type, or changing the composition of the medium. Modifications to apparatus may be useful, with proper justification and validation.
Many causes of variability can come from the formulation and manufacturing process. For example, poor content uniformity, process inconsistencies, a reaction taking place at different rates during dissolution, excipient interactions or interference, film coating, capsule shell aging, or hardening or softening of the dosage form as it ages may be sources of variability and interferences. During routine testing of the product, variability outside of the expected range should be investigated in terms of both analytical and formulation issues.